11.5.5 Concentrator Selection
There are three selections available for the Concentrator:
None . . . . . . . . . . . . . . . . . . . . . . . . Select this when no Concentrator will be
used with this Method.
Carbon. . . . . . . . . . . . . . . . . . . . . . . Select this when a Tri-Bed Concentrator will
be used with this Method.
Tenax . . . . . . . . . . . . . . . . . . . . . . . . Select this when a Tenax Concentrator will
be used with this Method.
11.5.6 Headspace Flow Parameter
CAUTION
Only trained users should modify methods. Incorrect
flow rates may lead to improper analysis.
Headspace Flow Pressure controls the flow rate of Nitrogen through the HSS
during the Sample and Purge states. This parameter is only available when the
HSS is used.
11.5.7 SituProbe Flow Parameter
CAUTION
Only trained users should modify methods. Incorrect
flow rates may lead to improper analysis.
SituProbe Flow Pressure controls the flow rate of Nitrogen through the SituProbe
during the Sample and Purge states. This parameter is only available when the
SituProbe is used.