5 - 2

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

5.2 Survey Mode

The Survey Mode is used for quick analysis and tentative results. The run period is
determined by the user, but it is generally two minutes long.

Sampling Technique

Sample the air away from the area of concern for one minute to determine the
background’s composition.

CAUTION

Do not touch the sample with the probe. Do not allow
liquids to enter the probe.

When the background has been determined, sample directly over the area of
concern to see what additional chemicals are present.

NOTE: See

Chapter 6, Methods

, for additional information on the Survey

method.

Procedure

1

Open the ER IQ software by double-clicking on the icon shown
in

Figure 5-1

.

Figure 5-1 ER IQ Icon

2

If the hand probe is attached, the HAPSITE will begin preparing to run a Survey
method. If this method is not the desired method, see

Section 5.3, Selecting a

New Method, on page 5-7

.

NOTE: The following message will appear during the power on sequence, see

Figure 5-2

.

Figure 5-2 NEG Full Power

Advertising