5 tune status icon – INFICON HAPSITE ER Chemical Identification System User Manual

Page 138

Advertising
background image

3 - 54

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

3.7.1.5 TUNE STATUS Icon

This icon provides information about the state of the HAPSITE’s tune. If the TUNE
icon is green, the tune status is good. The TUNE icon will turn yellow when a tune
check will be run in the near future. The icon will turn blue when the HAPSITE is in
the process of tuning. If the TUNE icon is red, a tune check needs to be run. If an
AutoTune fails, the AutoTune will turn gray and a FAIL message will be displayed.

To run a tune check, either touch PREPARE on the main screen and the HAPSITE
will run a tune check as part of its preparation, or press the RUN AUTOTUNE
button, which is shown at the bottom of the screen. This button allows for an
AutoTune to be run from the front panel. See

Figure 3-93

.

Figure 3-92 Run AutoTune

Touching the TUNE STATUS icon will also access the file name of the last tune
report, the time the instrument tuned, and the date the instrument tuned. It will also
show a countdown of the time to the next tune check.

The pressure of the MS at the last tune check will be displayed. If a method is not
running and the last tune pressure was within range, an Acceptable message will
be displayed. If a method is being run or manual tune is open, the current MS
pressure will also appear.

Advertising