15.8.1 Using ER IQ Software to Detach

1

Be sure to connect the HAPSITE to the Laptop using a crossover cable before
continuing. Refer to

section 2.2.5, Connect Laptop (if desired), on page 2-10

if

a connection needs to be made.

2

In ER IQ, left-click on the HAPSITE sensor to detach, then double-click the
Service Module icon, shown below in

Figure 15-23

.

Figure 15-23 Service Module Icon

3

Double-clicking on the Service Module icon will open the Service Module tab
in the HAPSITE Properties window, as shown in

Figure 15-24

.

Figure 15-24 Service Module Tab Selected in HAPSITE Properties Window

4

In the Service Module tab shown above, select the Detach button. A message
will appear looking for confirmation to detach the HAPSITE, as shown in

Figure

15-25

. Select Yes. A Detaching Service Module message is displayed. See

Figure 15-26

.

Figure 15-25 Detach HAPSITE Confirm Window