16.7.2 Part 1: NEG Pump Removal
1
This procedure requires the use of a Service Module. The HAPSITE should be
physically on the Service Module and attached. Refer to
for more information on
setting up and using the Service Module.
2
If the NEG Pump is hot from recent use, let the system cool overnight before
performing this procedure. The NEG Pump should cool to below 60
o
C before
proceeding.
WARNING
Burn Hazard. Do not bypass the cool down step! If NEG
is not cool upon removal, the NEG will become extremely
hot and possibly cause sparks.
3
Vent the HAPSITE by choosing Vent HAPSITE on the Service window of the
Service Module tab in the Properties window (see
). This starts an
automatic process which includes monitoring the cooling of the NEG Pump.
Figure 16-15 Vent HAPSITE from ER IQ Software