HAPSITE ER Operating Manual

16.7.2 Part 1: NEG Pump Removal

1

This procedure requires the use of a Service Module. The HAPSITE should be
physically on the Service Module and attached. Refer to

section 2.6, Service

Module, on page 2-17

and

Chapter 15, Service Module

for more information on

setting up and using the Service Module.

2

If the NEG Pump is hot from recent use, let the system cool overnight before
performing this procedure. The NEG Pump should cool to below 60

o

C before

proceeding.

WARNING

Burn Hazard. Do not bypass the cool down step! If NEG
is not cool upon removal, the NEG will become extremely
hot and possibly cause sparks.

3

Vent the HAPSITE by choosing Vent HAPSITE on the Service window of the
Service Module tab in the Properties window (see

Figure 16-15

). This starts an

automatic process which includes monitoring the cooling of the NEG Pump.

Figure 16-15 Vent HAPSITE from ER IQ Software