INFICON HAPSITE ER Chemical Identification System User Manual
Page 254
7 - 28
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
5a
If yes, then the BPG does not need adjustment. Proceed to
Step 8
.
5b
If no, the BPG will need to be adjusted. Proceed to
Step 6
.
6
To raise the BPG, increase the value of the EM Voltage by 25 volts. To lower
the BPG, decrease the value of the EM Voltage by 25 volts.
7
Check the BPG value to see if the BPG is now in the range (0.4 to 0.6).
7a
If the Base Peak Gain is still out of range, repeat
Step 6
until the value comes
into range.
7b
If the Base Peak Gain is in range, continue with
Step 8
.
8
Check the Tune and Mass Calibration Status panel to see if any of the masses
are highlighted in red. A red box indicates the mass is High or Low. See
Figure
7-32
.
Figure 7-32 BPG Good with Mass 50 Low
8a
If all mass percentages are within limits, then the instrument is tuned. Proceed
to
section 7.6.2, Adjusting Resolution, on page 7-29
8b
If any mass percentages are High or Low (red boxes), then proceed to
Adjusting Resolution.