INFICON HAPSITE ER Chemical Identification System User Manual
Page 78
2 - 40
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
Amber . . . . . . . . . . . . . . . . . . . . . . . The battery is fully charged. The rate of
charge has been reduced to a maintenance
level. The battery can be left this way
indefinitely.
Red . . . . . . . . . . . . . . . . . . . . . . . . . The receptacle (or the battery, if one is
installed) has a problem. A flashing red light
indicates that communication with the battery
is unsuccessful.
Off . . . . . . . . . . . . . . . . . . . . . . . . . . The receptacle is ready to charge a battery. If
the indicator remains extinguished when a
battery is inserted, the battery is severely
depleted. In this case, leave the battery in the
receptacle and unplug the power cord.
Reconnect the power cord and the battery
will start to charge.