6 part 5: activation of the neg pump, Figure 16-40 – INFICON HAPSITE ER Chemical Identification System User Manual
Page 510
Advertising
16 - 34
IP
N 07
4-
47
1-
P1
B
HAPSITE ER Operating Manual
Figure 16-40 Short Tune Icon
15
section 16.7.6, Part 5: Activation of the NEG Pump, on page 16-34
16.7.6 Part 5: Activation of the NEG Pump
1
Click on the Service Module tab of the Properties menu.
2
Click on the Service button. See
Figure 16-41 Service Button on the Service Module Tab
3
Use the Activate NEG settings below as a guideline.
HAPSITE Pump Down Time (min) = 60
NEG Activation Power (Watts) = 30
Activation Time (min) = 480
NEG Run Temp (C) = 400
Cooldown Time (min) = 720
See