2 setting up a quantitative search – INFICON HAPSITE ER Chemical Identification System User Manual

Page 380

Advertising
background image

11 - 42

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

Reject Bin . . . . . . . . . . . . . . . . . . . . Rejects peaks with less than a set number of

scans.

Deconvolution Level . . . . . . . . . . . High/Medium/Low. The default is Medium.

As the level of deconvoltution increases, the
program separates peaks that are closer
together.

Peak Shape . . . . . . . . . . . . . . . . . . . The shape requirement forces all of the

deconvoluted peaks to have the same
shape. As the shape requirement increases,
the shape of the individual ions must be
nearly the same.

The Search Trigger Criteria section of the Search page determines when an
AMDIS search will be run. There are three choices, see

Figure 11-35

.

Figure 11-35 Search Trigger Criteria

Every ____ Scans. . . . . . . . . . . . . . The value typed into this box determines the

interval at which AMDIS will run a search.
The default value for this is 100 scans.

Every ____ min:sec . . . . . . . . . . . . The value typed into this box determines the

time interval in which AMDIS will run a
search.

End of Run . . . . . . . . . . . . . . . . . . . If this is selected, an AMDIS search will only

be conducted at the end of a run.

11.9.2 Setting Up a Quantitative Search

CAUTION

Only trained users should modify methods. Changing
parameters may result in incorrect data.

When setting up a quantitative method, a quantitative library needs to be created
using the Calibration function of the software. See

Chapter 12, Target Compound

Methods

.

Advertising