HAPSITE ER Operating Manual

7.2.1 Starting AutoTune from the Manual Tune Screen on the Laptop

Computer

1

Double-click on the ER IQ icon. See

Figure 7-2

.

Figure 7-2 ER IQ Icon

2

Double-click on the Tune icon. See

Figure 7-3

.

HINT: Only the Advanced Access level can run Manual Tune. See

section

8.8.1, Changing Access Levels, on page 8-18

.

Figure 7-3 Manual Tune Icon

3

Wait until the EM and Emission buttons in the Control Panel turn green. Click
the Short Tune icon. See

Figure 7-4

.

Figure 7-4 Short Tune Icon

CAUTION

Adjusting other parameters without proper training may
damage the instrument.

4

Allow the Short AutoTune to run to completion and click OK. Close the Manual
Tune after receiving the message Final Results are on the LCD. See

Figure

7-5

.

Figure 7-5 AutoTune Complete