INFICON HAPSITE ER Chemical Identification System User Manual

Page 210

Advertising
background image

5 - 16

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

Figure 5-20 Headspace Needle

6

When the HAPSITE has been turned on (refer to

section Chapter 3 on page

3-1

) or taken out of Extended Standby (refer to

section 3.8.1 on page 3-61

), the

HAPSITE will automatically start preparing a Headspace method. If the method
that the HAPSITE begins preparing is not the desired one, refer to

Section 5.3,

Selecting a New Method, on page 5-7

.

6a

If running a Headspace method that requires a Concentrator, the HAPSITE will
run a Concentrator Cleanout. Its progress will be shown by a bar graph. If the
Concentrator Cleanout is not successful, refer to

Section , , on page 3-29

.

7

The HAPSITE will begin preparing to run the default Headspace method. It will
heat all necessary components, check pressures and run an AutoTune if
necessary. Progress of the preparation will be shown as a bar graph on the
Laptop screen. If AutoTune fails, see

Section 7.4, AutoTune Failure, on page

7-6

. When the final results message is displayed, click OK. See

Figure 5-21

.

Figure 5-21 AutoTune Complete

8

Once all temperature zones have reached their setpoints, a prompt will appear
Press RUN to start method.

NOTE: The sample must be heated for 20 minutes prior to sampling.

9

Gently press the Headspace needle into the 40 mL vial that contains the
blank/sample. See

Figure 5-20

.