HAPSITE ER Operating Manual

Figure 2-22 System Setup Screen with Service Module

NOTE: If the Service Module icon does not appear, the communications may

need to be configured. See

section 8.5.3, Establishing Communication

with the Service Module, on page 8-14

, for information on how to set

up a COM port for communicating with the Service Module.

5

Click on the Vent icon. The dialog shown in

Figure 2-23

will appear to confirm

the venting of the Service Module.

Figure 2-23 Vent Service Module Confirm Window

6

Select Yes to vent the Service Module. A window will appear counting down a
fifteen-second delay while the vent procedure completes. Once completed,
remove the aluminum plug.

7

Clean any debris or dust from the Viton® o-ring using a lint-free wipe with
methanol. Wipe the top of the o-ring, following the contour to clean the entire
top exposed section. Avoid pushing dust or debris into the middle opening.
Cleaning will ensure a tight seal to the HAPSITE, allowing the system to
maintain vacuum. See

Figure 2-24

, which shows the o-ring being cleaned.

WARNING

Wear gloves and safety glasses when handling methanol.