INFICON HAPSITE ER Chemical Identification System User Manual

Page 151

Advertising
background image

3 - 67

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

9

Gently push the Headspace needle into the 40 mL vial containing the
blank/sample.

10

The screen will prompt Collecting Sample Now For. See

Figure 3-114

. The

Headspace is automatically collecting a sample. No action is required by the
user.

Figure 3-114 Collecting Sample

11

The next screen will prompt Sampling is Done. Again, no action is required by
the user. See

Figure 3-115

.

Figure 3-115 Sampling Done HSS

12

By touching the CMPD ID, a list of found compounds will appear. The CAS
number, the Fit and the retention time for each compound will also be
displayed. This screen will also state the TIC (Total Ion Count) Max, the current
TIC count and the time the method has been running.

NOTE: Touching a compound on the list will display its Synonym and Exposure

Limit information.