HAPSITE ER Operating Manual
10
Replace the chain clamp insuring that the nut points to the left. Use the
torque wrench and tighten to 70 inch/pounds or 8 Nm. See
Figure 16-29
.
Figure 16-29 Torquing Chain Clamp
CAUTION
Do not connect the black plug from the new NEG pump to
the black socket at this time. This will avoid any
accidental heating of the NEG. pump.