INFICON HAPSITE ER Chemical Identification System User Manual

Page 258

Advertising
background image

7 - 32

IP

N 07

4-

47

1-

P1

B

HAPSITE ER Operating Manual

3

After making Ion Energy adjustments:

3a

If the red boxes no longer appear, indicating the Actual Resolution is within
range, proceed to

Step 4

. See

Figure 7-36

.

3b

If the HAPSITE is failing to Tune, then repeat the Manual Tune procedure.
Proceed to

section 7.6.1, Adjusting Base Peak Gain, on page 7-25

. If this

procedure is repeated more than twice, contact INFICON.

Figure 7-36 Good Manual Tune

4

Save the Tune file. Click Save Tune button on the upper Toolbar. Refer to

Figure 7-36

and see

Figure 7-37

.

Figure 7-37 Save Tune Button

246 180 230

263 185 250

282

190 255

Table 7-1 Manual Adjustments Guidelines (continued)

MASS

IE low