HAPSITE ER Operating Manual

4

Connect the end of the SituProbe Transfer Line to the HAPSITE at the same
location where the probe is normally connected. When inserting, line up the red
dots on the communication line with the red dots on the plug. Insert the Valco
connector into the top of the HAPSITE as shown. Screw the connector into
place. Connect the end, which is labeled in white with This End to Accessory,
to the back of the SituProbe accessory (IPN 932-220-G1).

Figure 2-32 Plugging Transfer Line Into HAPSITE and SituProbe

Figure 2-33 Attaching Transfer Line to HAPSITE

CAUTION

Do not open the SituProbe in a contaminated area. It is
not sealed against moisture, debris or contamination
with the front door open.

5

Open the front panel door of the SituProbe accessory and insert a Nitrogen
canister into the canister opening.

WARNING

Nitrogen canisters and cylinders are under pressure.