HAPSITE ER Operating Manual

15.7.1 Attaching the HAPSITE to the Service Module using the

ER IQ Software

Be sure to connect the HAPSITE to the Laptop before continuing. Refer to

section

2.2.5, Connect Laptop (if desired), on page 2-10

, if connection instructions are

needed.

1

Make sure that the HAPSITE does not heat. On the HAPSITE screen, touch
STOP PREPARE or using the arrow keys, highlight STOP PREPARE and
push OK SEL. See

Figure 15-13

.

Figure 15-13 Stop Prepare Button

2

In ER IQ, left-click on the desired HAPSITE sensor, then double-click the
Service Module icon, shown in

Figure 15-14

.

Figure 15-14 Service Module Icon in System Setup View, after Selecting HAPSITE Sensor

3

Double-clicking on the Service Module icon will open the Service Module tab
in the HAPSITE Properties window, as shown in

Figure 15-15

.

Figure 15-15 Service Module Tab Selected in HAPSITE Properties Window

4

In the Service Module tab shown above, select the Attach button. A
confirmation window will open. Click Yes. See

Figure 15-16

.