Changing edge detection order, Using video probes – HEIDENHAIN IK 5494-2D User Manual

Page 103

Advertising
background image

81

4

Probes

Changing edge detection order

Edges are detected as the scan moves from light to dark, from dark to light or automatically when the first

edge is encountered. Edge detection order is selected

from the VED toolbar.

Double-clicking a probe cycles the edge detection order of the

probe as shown here.

The edge detection order is indicated by the colors of the two

halves of each video probe.

The edge detection order must be consistent with the scan direction. For example, if the scan direction

(from cleanest to noisiest) is from a dark area of the image to a light area, the edge detection order must be

set to dark to light or first edge. When the edge detection order is inconsistent with the scan direction, no

points will be acquired.

by probe arrows in the dark

Dark to light order is indicated

by probe arrows in the light

First edge is indicated by a

single color probe

Dark to light is consistent with

the scan direction: probe fires

Automatic first edge: probe fires

Light to dark is not consistent

with the scan direction: probe

does not fire

Using Video Probes