Calibrating offset

Pixel offsets of all lower magnifications must be calibrated against the highest magnification to ensure that

specimens remain in the center of the view when switching between magnifications. These calibrations

are performed as part of the overall initial setup of the video probe system, and should be performed under

the following circumstances:

• The first offset calibration is performed at the highest magnification to establish the calibration

magnification standard and time/date stamp the calibration.

• Offset calibrations at lower magnifications are performed after performing all pixel resolution

calibrations.

• Offset calibrations at lower magnifications are performed without adjusting the focus obtained

at the highest magnification.

To calibrate offset:

1 Click the Probe/Probe Library menu item to display the probe library, and then click the Offset tab.

Calibrating Offset