QC5200 Series User’s Guide

3 Click the Teach button to begin the calibra-

tion process. The pixel calibration procedure will

use a circle artifact that you provide as a measure-

ment standard. You will be asked to enter the

artifact diameter.

4 Enter the artifact diameter us-

ing the current unit of measure and

click Continue. In this example, a

1 mm diameter artifact was used.

You will be instructed to measure

the artifact using the circle probe.

5 Measure the circle artifact and click OK to complete the calibration.

The resulting resolution values will be entered into the Pixel Resolution fields.

6 Click OK in the Probe Library screen to complete the pixel calibration, or select the next magnification

to be calibrated.