QC5200 Series User’s Guide
3 Click the Teach button to begin the calibra-
tion process. The pixel calibration procedure will
use a circle artifact that you provide as a measure-
ment standard. You will be asked to enter the
artifact diameter.
4 Enter the artifact diameter us-
ing the current unit of measure and
click Continue. In this example, a
1 mm diameter artifact was used.
You will be instructed to measure
the artifact using the circle probe.
5 Measure the circle artifact and click OK to complete the calibration.
The resulting resolution values will be entered into the Pixel Resolution fields.
6 Click OK in the Probe Library screen to complete the pixel calibration, or select the next magnification
to be calibrated.