Supervisor Setup

Specifying the maximum form error for a valid pixel calibration

The pixel resolution must be calibrated for each magnification to ensure accurate measurement results.

These calibrations are performed as part of the overall initial setup of the video probe system and are

discussed in detail in Chapter 4: Probes

Chapter 4: Probes.

The pixel calibration relies on the measurement of a standard circle artifact. Accurate calibrations require

high quality circle artifacts. The maximum form error allowed for circle artifact measurements can be

specified to insure that the circle used for calibration meets a known minimum quality standard.

Enter the maximum allowable form error into the Maximum

form for successful qualification field.

Including probes and points in live video image captures

Clicking the Tools/Capture Image menu item creates a bitmap (bmp) file of the live video window image

and stores the file in the QC5000\Parts folder using a name that includes time and date information.

The captured image always includes the video camera field

of view, and can also include the current video probe and

any points that might have been acquired when the probe was

fired.

Check the Include tools and points... box to include the probe

and acquired point images.

The circle probe and acquired points are

shown in the captured image

Video Probe Setup