347

1

0

Supervisor Setup

NOTE

A more detailed discussion of the worm tool is provided in Chapter 4: Probes

Chapter 4: Probes.

When the scan-limit is enabled, the limit boundary can be configured to prevent scanning within a speci-

fied number of pixels from the edge of the field of view. Limiting worm tool scanning to the center of the

field of view eliminates the effects of lens-edge aberrations.

Enter the desired boundary distance into

the Max distance from edge... field.

Specifying pattern file names

Whenever the pattern tool is taught a new pattern, it can be saved automatically in the Parts folder using a

default filename, or can be saved in another folder using a user-specified filename.

The default file naming convention is:

Part name_Pattern (n).5pt

Where n is incremented for each new file.

Check the Prompt for pattern file name... box to specify a

pattern file name and storage location whenever a pattern

is taught to the pattern tool.

FOV spherical and NLEC corrections

Normally, measurements are made at or near the center of the optical system to avoid aberrations normally

present near lens edges. Spherical and NLEC corrections compensate for

these aberrations, but can only be enabled by OEMs and distributors after

performing specific calibration procedures.

NOTE

Contact your OEM or distributor for more information regarding these two correc-

tions.

CAUTION

Do not change the status of the Enable FOV spherical correction or Enable FOV

NLEC check boxes without specific instructions from your OEM or distributor.