1

0

Supervisor Setup

Calibrating the display of magnification value

The video window image magnification

is displayed in the bottom-right corner

of the QC5000 main screen.

The image magnification depends on the

optical characteristics of the system and

must be found by performing the mea-

surement procedure provided below.

Calibration procedure

Step 1

Verify that the X pixel size... field is set to 1.000

Step 2

Click Tools, Customize and then click

the Status bar tab. True magnification

must be enabled. If the True magnifica-

tion box is not checked, click on True

magnification and then click Show.

Step 3

Click OK to close the Customize screen, and then return to the Tools/Options/VED tab.