QC5200 Series User’s Guide

Add step: Goto selected feature

At this step in the programming process, the point returned from the center of the hole pattern is the cur-

rently selected feature. Adding a Goto Selected Feature step at this time will cause the system to position

the stage with the hole feature at the center of the field of view. This key step guarantees that the hole will

be located in a known position prior to its measurement, independent of part variability.

Measure feature

Now that the hole feature is located at the center of the field of

view, the Circle probe is used to measure it.

Default probe positioning is based on the assumption that part

variability is very low, so absolute probe coordinates are re-

corded as the program is created, and then used again to posi-

tion the probe each time the program is executed. When part

variability is an issue, the recorded probe position is no longer a

reliable indicator of the required probe position for new parts.

Since the last CNC move positioned the hole feature at the

center of the field of view, the required probe position is also

at the center of the field of view. Therefore, preventing any

Adding a Goto Selected Feature step...

positions the target hole

feature in the center of the

field of view

The hole is measured at the center of

the field of view