5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1001

Parameters
The parameters for the command MEAS T/Flow are configured on the
following 4 tabs:

General/Hardware
Parameters for device.

Measuring parameters
Parameters for setting the measurement procedure.

Evaluation
Defining further methods for evaluating measurement curves.

Additional measured values
Defining further measured values which can be saved as additional col-
umns in the measuring point list.

Command variables
The following command variables are generated in the method run by the
command MEAS T/Flow and can be used in formulas under the designa-
tion 'Command name.Variable designation':

Identifica-
tion

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has not ever been started

.DBL

Total duration for the processing of the command in s

.EGF

Last measured gas flow rate (measured value follow-
ing processing of the command) in mL/min

.EME

Final measured value (measured value after processing
of the command) in the unit of the measured value

.FIN

Command status; 1 = Command has been ended
at least once
, 0, invalid (variable not available) =
Command has never been ended

.GMA

Maximum gas flow rate in mL/min

.GMI

Minimum gas flow rate in mL/min

.GMN

Average gas flow rate in mL/min

.IGF

Initial gas flow (measured value at the time of the start
of the command) in mL/min