5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1104

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Reference temperature

Input of a temperature for which a calibration point of the conductivity
standard is known. It should be as close as possible to the temperature at
which measurements are carried out.

Range

-20.0 ... 150.0°C

Default value

20.0

Conductivity

Input of the conductivity of the standard at the reference temperature.

Range

0.00000 ... 2000.00 mS/cm

Default value

1.0 mS/cm

Temperature coefficient

Temperature coefficient of the standard with the reference temperature.

Range

0.00 ... 9.99 %/°C

Default value

1.00 %/°C

5.6.5.5

CAL Spec

5.6.5.5.1

CAL Spec - Overview

Dialog window: Method

CAL Spec Properties... CAL Spec - 'Command

name'

Command for wavelength calibration.

Instruments
This command can be executed with the following devices:

Spectrometer: Avantes

Appearance
The command has the following appearance:

Parameters
The parameters for the command CAL Cond are configured on the fol-
lowing 3 tabs:

General/Hardware
Parameters for device.

Measuring parameters
Parameters for setting the measurement procedure.