Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1101

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1085

855
Selection

1 | 2 | 3

Default value

1

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Note

If an existing solution is selected as the solution to be added then the
Concentration and Unit of this solution will be adopted for calculat-
ing the standard addition. If not defined is selected, then Concentra-
tion
and Unit must be entered on the Standard addition tab.

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando, 867
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Sensor

Selection of a sensor of type ISE electrode from the sensors configured
in the sensor table.