5 cal cond - data conductivity standard – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1119

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1103

Reference temperature

Input of a temperature for which a calibration point of the conductivity
standard is known. It should be as close as possible to the temperature at
which measurements are carried out.

Range

-20.0 ... 150.0°C

Default value

20.0

Conductivity

Input of the conductivity of the standard at the reference temperature.

Range

0.00000 ... 2000.00 mS/cm

Default value

11.66 mS/cm

Temperature coefficient

Temperature coefficient of the standard with the reference temperature.

Range

0.00 ... 9.99 %/°C

Default value

2.07 %/°C

Parameters for template
If the option Input has been selected, then the following parameters will
be displayed:

Name standard

List box with conductivity standards. The data of one of these standards is
used to calculate the temperature coefficient for the specified reference
temperature.

Selection

'Conductivity standard'

5.6.5.4.5

CAL Cond - Data Conductivity standard

Dialog window: Method

CAL Cond Data Conductivity standard

Current temperature

Display of the currently measured temperature of the conductivity stand-
ard.

Data transmission from standard
Input of the data of the conductivity standard, which should be as close as
possible to the Current temperature. This data is as a rule contained in
the Data Sheet/Certificate for the conductivity standard.