7.20 IO controller

■■■■■■■■■■■■■■■■■■■■■■

1632

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

'Port description'
Only analog output ports of the Controller output type can be selec-
ted.

Control task

Selection of the control task.

Selection

Pump control | Free controller

Default value

Pump control

Note

Pumps with analog input should be used for the dosing performance in
the presence of pump control.

Control parameters

Controller amplification

Amplification of the controller.

for control task = pump control
Range

0 ... 214783647.000

Default value

1

for control task = free controller
Range

0 ... 214783647.000

Default value

0

P in feedback path

on | off (Default value: off)

Reset time

Time which defines the gradient of the rise of a constant input.

for control task = pump control
Range

0.0 ... 9999.0

Default value

0.0

for control task = free controller
Range

0.0 ... 9999.0 s

Default value

0.0 s

Rate time

Time that a P controller requires, with constant modification speed of the
controlled variable (ramp function), to bring about the same modification