Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 848

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

832

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

851, 852
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will be used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.

Cell

Generator type

Selection of the generator type.

Selection

without diaphragm | with diaphragm

Default value

without diaphragm

Generator current

Polarization current at the generator electrode.

for generator type = without diaphragm
Selection

100 mA | 200 mA | 400 mA | auto

Default value

400 mA

auto
Means that the current is automatically adapted to the conductivity of
the reagent and that near the endpoint the current will be controlled
at smaller values.

for generator type = with diaphragm
Selection

100 mA | 200 mA | 400 mA | auto

Default value

auto

auto
Means that the current is automatically adapted to the conductivity of
the reagent and that near the endpoint the current will be controlled
at smaller values.