5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1189

Note

The dosing rate should be reduced for viscous liquids.

Dosing time

Time during which dosing is to be carried out. This parameter is only dis-
played for Dosing criterion = Volume/Dosing time or Dosing rate/
Dosing time.

Range

0 ... 9999999 s

Default value

100 s

Pause

Waiting period before dosing is started.

Range

0 ... 999999 s

Default value

0 s

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously. This value is used for
temperature correction during monitoring with pH measurements..

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando, 855
Range

0.1 ... 999999 s

Default value

5.0 s

Titrino
Range

1 ... 999999 s

Default value

5 s