2 result - monitoring, Monitoring – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1259

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1243

Description

Freely selectable entry of a description of the result.

Input

1024 characters

[Save as template]

Open the dialog window Save result template in order to save the
result parameters as template for the creation of new results (see Chapter
5.6.8.2.6.2, page 1247)
.

5.6.8.2.5.2

Result - Monitoring

Tab: Method

CALC Properties... [New] [Continue] Result - Overview

The parameters for monitoring the result are defined on this tab.

Result monitoring

Result monitoring

on | off (Default value: off)

The question of whether the result is within the defined limit values is
monitored while calculating the result during the determination if this
option is enabled.

Lower limit

Lower limit for the result.

Input

10 digits

Upper limit

Upper limit for the result.

Input

10 digits

Message

Message

The message defined here can be put out to different targets if the upper
limit is exceeded or if the lower limit is not reached. The text editor for

entering or changing the message is opened with

or by double-click-

ing into the text field (see Chapter 2.5.2, page 93). The formula editor can
also be called up inside the text field (see Chapter 2.4, page 32).

Input

Text (unlimited)