6 Configuration

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1425

Calibration date

Date and time of last calibration that is entered automatically after each
automatic calibration or manual entry (is not being displayed for the Ther-
moprobe).

Calibration method

Name of the method with which the last calibration was carried out (is
not being displayed for the Thermoprobe). If the calibration data has been
entered manually, then manual will appear here.

User

Short name of the user logged in during the calibration or when the cali-
bration data was entered manually (is not being displayed for the Ther-
moprobe). If work is not carried out using login, then the user logged in
under Windows will be entered automatically.

Measuring input

Display of the type and the serial number of the measuring input at which
the sensor was connected during the calibration (only for Titrando, is not
being displayed for the Thermoprobe).

[Initial data]

Open the dialog window Initial data sensor (see Chapter 6.7.3.7, page
1434)
, in which the initial calibration data stored in the sensor chip is dis-
played (only for intelligent sensors).

Monitoring the calibration data
on | off (Default value: off)

If this option is activated, then the validity of the calibration will be moni-
tored (is not being displayed for the Thermoprobe).

Calibration interval

Validity of the calibration in days. If a value is entered here then the date
for the Next calibration will be adapted automatically.

Note

The days are monitored, e.g. the calibration becomes invalid when the
date changes and not at the time time of day that calibration took
place.

Can only be edited if monitoring is switched on.
Range

0 ... 999 Days

Default value

999 Days