5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

910

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Titrando, 855
Range

0.0 ... 150.0 °C

Default value

0.2 °C

Action

Selection of the event to be triggered if the upper or lower limit is excee-
ded:

Selection

Cancel determination | Cancel command
(Titrando, 855)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT command will be canceled, then the exit track (if
present) will be started and the determination will be finished.
Cancel command (Titrando, 855)
The running STAT command will be canceled, then the next com-
mand will be carried out.
Wait for [Continue]
Reagent dosing in the running STAT command will be interrupted and
a message will be displayed. As soon as the monitored measured value
is again within the limits (including hysteresis) reagent dosing can be
resumed by pressing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current STAT command will be interrupted. As
soon as the monitored measured value is again within the limits
(including hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.

Track call on limit exceeding
In this table, which cannot be edited directly, a maximum of 20 entries
can be defined as to which track is to be started automatically when a
particular limit value is exceeded.

[New]

Open the dialog window Track call # in which the parameters for the
track call can be entered (see Chapter 5.6.3.8.4, page 861).

[Properties]

Open the dialog window Track call # in which the parameters for the
track call can be edited (see Chapter 5.6.3.8.4, page 861).

[Delete]

Delete the track call selected in the table.