5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1176

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Monitoring Measured value

Monitoring Measured value

on | off (Default value: off)

If this option is switched on then the measured value will be monitored
and any exceeded limits will be entered in the measuring point list.

Lower limit pH

Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is trig-
gered.

Titrando, 855
Range

-20.000 ... 20.000

Default value

-20.000

Titrino
Range

-20.00 ... 20.00

Default value

-20.00

Lower hysteresis pH

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value
lower limit ok
is triggered.

Titrando, 855
Range

0 ... 20.000

Default value

20

Upper limit pH

Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit exceeded is trig-
gered.

Titrando, 855
Range

-20.000 ... 20.000

Default value

20.000

Titrino
Range

-20.00 ... 20.00

Default value

20.00