5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

869

855
Selection

1 | 2 | 3

Default value

2

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method run to see whether the
correct solution has been set on the selected dosing device and whether
the dosing device type is correct. With non-intelligent exchange or dosing
units, only the cylinder volume is checked.

Titrando, 855
Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Filling rate

Rate at which the dosing cylinder of the second dosing device is to be
refilled. The maximum filling rate depends on the cylinder volume of the
exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device, it will automatically be reduced during dos-
ing to the largest possible value (see Chapter 5.6.3.8.5, page 862).

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1