5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

961

[New]

Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 932)
.

[Properties]

Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 932)
.

[Delete]

Delete the external measured value selected in the table.

5.6.4.4.6

MEAS U - Fixed endpoint evaluation #

Dialog window: Method

MEAS U Properties... Additional evalua-

tions

Fixed endpoint evaluation [New]/[Property] Fixed endpoint evalua-

tion #

Measured quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value (Titrando, 867, 855)
Range

-2000.0 ... 2000.0 mV

Measured value (Titrino)
Range

-2000 ... 2000 mV

Time
Range

0.0 ... 999999.9 s

5.6.4.5

MEAS Ipol

5.6.4.5.1

MEAS Ipol - Overview

Dialog window: Method

MEAS Ipol Properties... MEAS Ipol - 'Command

name'

Command for voltametric measurements with selectable polarization
current.