5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1158

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

off
Means that no stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Titrando, 846, 856, 867, 814, 815, 855, 864, 874
Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this option is enabled, the stirrer will be switched off automatically
when the dosing has finished. This parameter is displayed only for Titran-
dos, 846, 856, 867, 814, 815, 855, 864 and 874.

5.6.6.2.3

ADD - Dosing parameters

Tab: Method

ADD Properties... Dosing parameters

Command name

Name of the command.

Input

25 characters

The volume to be added and the dosing and filling rate are defined on this
tab.

Volume

Fixed volume that is to be dosed.

Titrando, 846, 814, 815, 855, 864, 874
Range

0.00000 ... 99999.9 mL

Default value

10.0000 mL

Titrino
Range

0.0001 ... 9999.0 mL

Default value

10.0000 mL

730, 778, 789
Range

0.001 ... 999.999 mL

Default value

10.0000 mL

774
Range

0.01 ... 999.999 mL

Default value

10.0000 mL