4 properties - tower, Properties - tower # 7 – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1573

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

7 Instruments

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1557

Measuring input 1/2 (iConnect)

ADC type

Shows the type of analog-digital converter.

Serial number

Shows the serial number of the measuring input interface.

Temperature sensor

Selects the type of temperature sensor connected to the measuring input.

With the 856 Conductivity Module displayed only.
Selection

Pt 1000 | NTC

Default value

Pt 1000

R (25 °C)

Nominal resistance of connected NTC sensor.

Only visible with temperature sensors of the NTC type.
Range

1000 ... 99999 Ohm

Default value

30000 Ohm

B value

Material constant of the NTC resistance referred to measuring the resist-
ance at 25 °C and 50 °C.

Only visible for temperature sensors of the NTC type.
Range

1000 ... 9999

Default value

4100

7.12.4

Properties - Tower #

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Tower #

On the tab Tower the properties of the tower and of the robotic arm
connected to are displayed.

Tower parameters

Max. stroke path

Entry of the lowest permitted lift position. A lift height of 0 mm corre-
sponds to the "Home position", i.e. the lift is moved right to the top.

Range

0 ... 235 mm

Default value

235 mm