5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

775

5.6.3.4.6.5

SET Upol - Titration parameters

Tab: Method

SET Upol Properties... Titration parameters

Command name

Name of the command.

Input

25 characters

Parameters defining the run of the titration.

Titration direction

Selection of the titration direction. If two endpoints are set, then the titra-
tion direction is already established and the following selection of titra-
tion direction
is ignored.

Selection

+ | - | auto

Default value

auto

+
positive measured value alteration
-
negative measured value alteration
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.

Extraction time

The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). For example, it makes sense to
enter an extraction time for the titration of samples that are not easily
soluble.

Titrando, 719, 720, 736, 751, 758, 785, 794, 798, 799
Range

0 ... 999999 s

Temperature

Manually entered titration temperature. If a temperature sensor is connec-
ted and the Temperature measurement is set on the General/Hard-
ware
tab under Sensor to automatic or continuous, then the temper-
ature will be measured continuously. This value is used for temperature
correction in pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C