5.6 Commands
■■■■■■■■■■■■■■■■■■■■■■
1144
■■■■■■■■
tiamo 2.3 Patch 1 (for Process Analysis)
Temp. (°C)
Bec4
Bec7
Bec10
pH 4
pH 7
pH 10
55
4.08
6.98
-
60
4.09
6.98
-
65
4.11
6.99
-
70
4.12
6.99
-
75
4.14
7.00
-
80
4.16
7.00
-
85
4.18
7.01
-
90
4.19
7.02
-
95
4.21
7.03
-
Radiometer
Temp. (°C)
Rad4.01
Rad7.00
Rad9.18
pH 4.01
pH 7
pH 9.18
0
4.000
7.118
9.464
5
3.998
7.087
9.395
10
3.997
7.059
9.332
15
3.998
7.036
9.276
20
4.001
7.016
9.225
25
4.005
7.000
9.180
30
4.011
6.987
9.139
35
4.018
6.977
9.102
40
4.027
6.970
9.068
45
4.038
6.965
9.038
50
4.050
6.964
9.011
55
4.064
6.965
8.985
60
4.080
6.968
8.962
65
4.097
6.974
8.941
70
4.116
6.982
8.921
75
4.137
6.992
8.900