8 Manual control

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1667

702, 716, 718, 719, 720, 721, 736, 794
Selection

Bit-pattern with exactly 8 characters (0, 1, *,
p)
| ******** | Signal pattern

Default value

********

Spectrometer
Selection

Bit-pattern with exactly 10 characters (0, 1, *,
p)
| ********** | Signal pattern

Default value

**********

The output lines are numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0

Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0

Examples:

************1* sets the output line 1 to enabled (= set), which would
result in a stop command for a connected Titrino, for example.

************0* sets the line to inactive.

Note

We recommend masking the irrelevant output lines with an asterisk *
so as not to modify these line conditions.

Note

For Titrinos with 3 output lines, only 3 characters can be entered. If sig-
nal pattern is selected, also only the first 3 characters will be used.

[Set]

Set the bit pattern defined under Outputs.

8.4.4

Sample changer functions

8.4.4.1

Sample changer - Overview

Program part: Manual control

If in the subwindow for the device selection, a tower of a sample changer
is selected, then the functions and parameters belonging to this sample
changer are displayed in the subwindow Functions/Parameters.

Sample changer functions
The sample changer functions are displayed on the following tabs: