Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 604

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

588

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

[Delete]

Delete the selected line.

pK/HNP evaluation

pK/HNP evaluation

on | off (Default value: off)

If this option is enabled, then the pK value is determined from the titration
curve which corresponds to the pH value at the half neutralization point
(see Chapter 5.6.3.10.1, page 927).

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the meas-
uring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 20.0 pH/mL

Default value

1.0 pH/mL

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the meas-
uring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

0.1 ... 20.0 pH/mL

Default value

1.0 pH/mL