6 Configuration

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1399

Selection

mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | ppm | % | mEq/L

Default value

mol/L

Date titer det.

Display of the date and time of last titer determination, which is automati-
cally entered with each automatic titer determination or manual entry.

Titer method

Name of the method with which the last titer determination was carried
out. If the titer has been entered manually then manual will be shown
here.

User

Short name of the user logged in during titer determination or when the
titer was entered manually. If work is not carried out using login, then the
user logged in under Windows will be entered automatically.

Statistics
Statistical data will only be shown if the titer has been assigned by the
titer method as a mean value of a multiple determination. If the titer has
been determined with a single determination or entered manually then no
statistics values will be shown.

Number of titer determinations (n)

Number of determinations used to determine the titer.

Abs. standard dev. (s abs)

Absolute standard deviation for the titer determination in the units of the
titer.

Rel. standard dev. (s rel)

Relative standard deviation for the titer determination in %.

Monitoring the titer validity
on | off (Default value: off)

If this option is activated, then the validity of the titer will be monitored.

Validity

Validity of the titer in days. If a value is entered here then the date for the
Next titer determination will be automatically adapted.

Range

0 ... 999 Days

Default value

999 Days