5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

681

5.6.3.3.6.3

MET Cond - Start conditions

Tab: Method

MET Cond Properties… Start conditions

Command name

Name of the command.

Input

25 characters

The start conditions are processed in the listed sequence before the con-
ductivity measurement is started.

Initial measured value
The initial measured value is determined before the other start conditions
are processed. The measured value acceptance depends on the three fol-
lowing parameters:

Signal drift

The measured value is only accepted if the drift is less than the value
entered here.

Range

0.001 ... 999.0 (mS/cm)/min

Selection

off

Default value

off

off
The measured values will be accepted when the maximum waiting
time has elapsed.

Min. waiting time

The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.

Range

0 ... 999999 s

Default value

1 s