5 brc - titration parameters, Titration parameters – Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 869

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

853

Range

1.0 ... 999.0 µg/min

Default value

15.0 (slow), 25.0 (optimal), 50.0 (fast) µg/min

Selection

minimum

Stop criterion

Stop criterion

Switches off titration when the endpoint has been reached and the stop
criterion has been met.

Selection

Drift | rel. drift | Time | drift & time | off

Default value

drift & time

Stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift or drift
& time
.

Range

1 ... 999 µg/min

Default value

15 µg/min

Relative stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion
= rel. drift.

Range

0 ... 999 µg/min

Default value

5 µg/min

Delay time

If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time or drift & time.

Range

0 ... 999 s

Default value

40 s

5.6.3.7.5

BRC - Titration parameters

Tab: Method

BRC Properties... Titration parameters

Command name

Name of the command.

Input

25 characters

The parameters for the running of the titration are defined on this tab.