6 Configuration

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1405

Range

1 ... 9

Default value

2

Dosing rate

Rate at which the solution is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit used (see below).
When the function is carried out the rate will automatically be decreased
to the highest possible value.

Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Filling rate

Rate at which the cylinder is to be filled. The maximum filling rate depends
on the cylinder volume of the exchange unit used (see below). When the
function is carried out the rate will automatically be decreased to the
highest possible value.

Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Maximum dosing/filling rate for the exchange unit, depending on
the cylinder volume:

Cylinder volume

maximum rate

1 mL

3.0 mL/min

5 mL

15.0 mL/min

10 mL

30.0 mL/min

20 mL

60.0 mL/min

50 mL

150.0 mL/min

Note

Enter lower rates for high-viscosity liquids.

Tubing parameters
Definition of the length and diameter of the tubing connected to the
exchange unit.