5 Method

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1205

Identifi-
cation

Description

.FIN

Command status; 1 = Command has been ended at
least once
, 0, invalid (variable not available) = Com-
mand has never been ended

5.6.6.6.2

EMPTY - Properties

Dialog window: Method

EMPTY Properties... EMPTY - 'Command name'

Command name

Name of the command.

Input

25 characters

Device

Device name

Selection of a device from those available in the device table. The only
devices which shall be offered are those with which the command can
run.

Selection

'Device name' | not defined

Default value

not defined

not defined
In this case, the device name must be assigned by the user at the start
of the method.

Device type

Display or selection of the device type. If a device is selected under Device
name
, then this Device type field can no longer be edited, and the
device type belonging to the selected device is displayed instead

If the option not defined is selected as Device name, any device types
or device groups which are able to execute the command can be selected,
independently of the devices in the device table.

Selection

Device types | Titrando

Default value

Titrando

Dosing device

Dosing device

Selection of the dosing device (Dosinos only) to be emptied. All the dosing
device connections which are possible with the selected device type are
displayed.