Metrohm tiamo 2.3 Patch 1 (process analysis) User Manual

Page 1703

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

8 Manual control

tiamo 2.3 Patch 1 (for Process Analysis)

■■■■■■■■

1687

Start delay

Waiting time in seconds or minutes before the initiated action is started.

Range

0 ... 999999

Default value

0

Selection

s | min

Default value

s

Working time

Time in seconds or minutes for which the digital output is set to ACTIVE.
After the working time has elapsed, it is set to INACTIVE again.

Range

0 ... 999999

Default value

0.01

Selection

s | min

Default value

s

End delay

Waiting time after the working time has expired

Range

0 ... 999999

Default value

0

Selection

s | min

Default value

s

Status

Display of the current status.

Selection

ACTIVE | INACTIVE | ERROR

Time to finish

Display of the time to finish of the timer control.

[Start]

Sets the digital output to ACTIVE.

[Stop]

Sets the digital output to INACTIVE.

Setting a default

Port description

Display of the port description.

Port

Display of the internal description of the port.