5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1206

■■■■■■■■

tiamo 2.3 Patch 1 (for Process Analysis)

Titrando, 846, 856, 867
Selection

1 | 2 | 3 | 4

Default value

1

814, 815, 855, 864, 874
Selection

1 | 2 | 3

Default value

1

730, 774
Selection

1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12

Default value

1

Titrino
Selection

external D1 | external D2

Default value

external D1

Solution

Input of a solution name or selection of a solution from the solutions
defined in the solution table. If intelligent exchange or dosing units are
used, then a check will be made in the method sequence to see whether
the correct solution has been set on the selected dosing device and
whether the dosing device type is correct. With non-intelligent exchange
or dosing units, only the cylinder volume is checked. At the start of the
command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

5.6.6.7

RLS DOS

5.6.6.7.1

RLS DOS - Overview

Dialog window: Method

RLS DOS Properties... RLS DOS - 'Command name'

Command to release a dosing device for using it in other parallel run-
ning methods. It removes the reservation made by the current method.

Devices
This command can be executed with the following devices:

Sample Processor: 730, 774, 778, 789, 814, 815, 864, 874